抽象的

Amperometric determination of copper (II) in alloy samples and synthetic mixtures using gallacetophenone phenylhydrazone

D.Venkataramana Reddy, M.Jagadeesh, A.Varada Reddy


Gallacetophenone phenylhydrazone (GPPH) has been used as an analytical reagent for amperometric determination of copper. Copper(II) is quantitatively determined byGPPH at pH 4.0-6.0.After studying the polarographic behaviour of GPPH and copper(II) at dropping mercury electrode (DME), applied potential is fixed at -0.4Vvs saturated calomel electrode (SCE). The method is applied for the determination of copper in brass, chalcopyrites and syntheticmixtures. The stoichiometry of the complex is found to be 3:2 (metal : ligand). This is also confirmed fromthe micro analytical data of the solid complex, thermo gravimetric and differential thermal analysis curves of the complex.


索引于

  • 中国社会科学院
  • 谷歌学术
  • 打开 J 门
  • 中国知网(CNKI)
  • 引用因子
  • 宇宙IF
  • 米亚尔
  • 秘密搜索引擎实验室
  • 欧洲酒吧
  • 巴塞罗那大学
  • ICMJE

查看更多

期刊国际标准号

期刊 h 指数

Flyer