抽象的

Development and validation of headspace method for determination of residual solvents in diphenoxylate hydrochloride bulk drug

R.S.Lokhande, P.U.Singare, P.V.Jadhav


Organic volatile impurities by Headspace GC method was developed and validated for Diphenoxylate hydrochloride bulk drug for its seven residual solvents.Analysis was performed on Perkin Elmer HS 40 system with auto injector. Carrier gasHeliumwas used with constant flowrate of 2.5mL/min as carrier gas and the separation of residual solvents were achieved on DB- 5 column. The thermostat temperaturewas 125 °C for 20 minute for each vial and after the equilibration the vials were pressurized and injected on GC column. FID detector was used for detection. The parameter for which the method was validated included specificity, limit of detection and quantification, linearity, precision, accuracy and robustness. The method was successfully used to quantify the levels of specified limit for residual solvents in Diphenoxylate hydrochloride bulk drug.


索引于

  • 中国社会科学院
  • 谷歌学术
  • 打开 J 门
  • 中国知网(CNKI)
  • 引用因子
  • 宇宙IF
  • 电子期刊图书馆
  • 研究期刊索引目录 (DRJI)
  • 秘密搜索引擎实验室
  • ICMJE

查看更多

期刊国际标准号

期刊 h 指数

Flyer