抽象的

Development of a Validated Stability Indicating RP-HPLC Method for Simultaneous Determination of Benazepril, Benazeprilat, and Amlodipine in Presence of Their Degradation Products and Application to Pharmaceutical Formulation and Human Plasma

Hemdan A, Omar AE, Farouk M, Shereen MT and Mostafa AS


A sensitive, reproducible, and rapid stability indicating RP-HPLC method was developed and subsequently validated for simultaneous determination of both benazepril (BENZ), benazeprilat (BENZT), and amlodipine (AML) in presence of their basic hydrolysis degradation products in bulk powder, pharmaceutical formulation, and application to human plasma using moexipril (MOX) as an internal standard (IS). The method uses Inertsil C18 column (250 mm × 4.6 mm, 5 μ) and the mobile phase used was consisting of acetonitrile-potassium hydrogen phosphate buffer pH 7 adjusted by 0.1 N NaOH (40/60 v/v) with a flow rate of 1.0 mL/min and detection at 242 nm. A detailed validation of the methods was performed following the ICH guidelines and the standard curves were found to be linear in the range of 0.5 μg/mL to 100 μg/mL, 0.5 μg/mL to 100 μg/mL and 5 μg/mL to 100 μg/mL for BENZ, BENZT, and AML, respectively. Statistical comparison was done between the proposed method and the reported one where no significant difference was found between the two methods.


索引于

  • 中国社会科学院
  • 谷歌学术
  • 打开 J 门
  • 中国知网(CNKI)
  • 引用因子
  • 宇宙IF
  • 电子期刊图书馆
  • 研究期刊索引目录 (DRJI)
  • 秘密搜索引擎实验室
  • ICMJE

查看更多

期刊国际标准号

期刊 h 指数

Flyer