抽象的

Differential pulse polarographic determination of ultra trace nickel in natural waters after solid phase extraction

Mahmood Payehghadr, Mahsa Mirzaei


A highly selective and sensitive anodic stripping differential pulse polarographic method for determination of ultra trace amounts of Ni2+ ions in water samples is proposed. The method is based on the separation and preconcentration of Ni2+ on an octadecyl bonded silica membrane disk modified by a recently synthesized Schiff base, 2-((E)-(2-(2-(pyridine-2-yl) ethylthio) ethylimino) methyl)-4-bromophenol (PYTABr) at pH=7. The retained nickel on the membrane was eluted with 10 ml 0.1 M HNO3 and determined by anodic stripping differential pulse polarography. Extraction efficiency, influence of sample volume and eluent flow rates, amount of ligand, amount of eluent for elution of Ni2+ ions from membrane disk, pH, breakthrough volume and limit of detection have been evaluated. The effect of foreign ions on the percent recovery of heavy metal ions has also been studied. The limit of detection of proposed method for Ni2+ was found to be 3.2ng. L-1 and also an enrichment factor of 200were obtained. Themaximum capacity of the membrane disksmodified by 5 mg of PYTABr was found to be 325± µg Ni2+. The proposed method has been applied to the determination of nickel in several water samples with satisfactory results.


免责声明: 此摘要通过人工智能工具翻译,尚未经过审核或验证

索引于

  • 中国社会科学院
  • 谷歌学术
  • 打开 J 门
  • 中国知网(CNKI)
  • 引用因子
  • 宇宙IF
  • 电子期刊图书馆
  • 研究期刊索引目录 (DRJI)
  • 秘密搜索引擎实验室
  • ICMJE

查看更多

期刊国际标准号

期刊 h 指数

Flyer