抽象的

RP-HPLC Estimation of Tadalafil in Tablet Dosage Form

A. S. Sutar, C. S. Magdum, S. S. Patil and N. S. Naikawadi


A simple, specific, accurate and precise method, reverse phase high performance liquid chromatography was developed for estimation of tadalafil in tablet dosage form. For HPLC method, Hi- Qsil C18-10 column in isocratic mode with mobile phase containing acetonirtile; acetate buffer pH (2.8) in ratio of 45 : 55 v/v was used. The flow rate was 1 mL/min and effluent was monitored at 283 nm. Retention time was found to be 4.46 ± 0.03 minutes. Method was validated in terms of linearity, accuracy and precision. Results of the analysis show that the method can be used for the estimation of tadalafil in tablet dosage form.


索引于

  • 中国社会科学院
  • 谷歌学术
  • 打开 J 门
  • 中国知网(CNKI)
  • 引用因子
  • 宇宙IF
  • 米亚尔
  • 秘密搜索引擎实验室
  • 欧洲酒吧
  • 巴塞罗那大学
  • ICMJE

查看更多

期刊国际标准号

期刊 h 指数

Flyer