抽象的

SimultaneousRP-HPLCdetermination of bambuterolHCl and montelukast sodiumin pharmaceutical preparations

V.V.Vaidya, G.R.Singh, M.B.Kekare, M.P.Choukekar


A simple, fast and precise reversed phase high performance liquid chromatographic method is developed for the simultaneous determination of bambuterol HCl and montelukast sodiumusing mebeverine hcl as an internal standard. Chromatographic separation was performed on a waters symmetry C18 column (150mm4.6mm, 5m) as stationary phase with amobile phase comprising of 0.05% trifluoro acetic acid in water: 0.05% trifluoro acetic acid in acetonitrile (45:55 v/v), at a flow rate of 0.7mL min-1 and UV detection at 215nm. The Retention time of Bambuterol hcl, montelukast sodiumand Mebeverine HCl were 1.270 min, 4.322 and 1.604 min respectively. The proposed method was validated for linearity, accuracy, precision, LOD, LOQ. Linearity, accuracy and precisionwere found to be acceptable over the ranges of 250-750g mL-1 for both bambuterol HCl and montelukast sodium. It can be conveniently adopted for routine quality control analysis.


索引于

  • 中国社会科学院
  • 谷歌学术
  • 打开 J 门
  • 中国知网(CNKI)
  • 引用因子
  • 宇宙IF
  • 电子期刊图书馆
  • 研究期刊索引目录 (DRJI)
  • 秘密搜索引擎实验室
  • ICMJE

查看更多

期刊国际标准号

期刊 h 指数

Flyer