抽象的

Solid phase extraction of ultra trace amounts of cobalt and nickel on octadecyl bonded silicamembrane diskmodifiedwithCyanex272 and determination by inductively coupled plasma-optical emission spectrometry

Ali Jafari, Ali Moghimi


A highly sensitive and accurate method for preconcentration and determination of ultra trace amounts of cobalt and nickel ions in water samples is proposed. The preconcentration is achieved using C18-silica extraction disksmodifiedwithCyanex272(bis(2,4,4-trimethyl pentyl) phosphinic acid). The retained ions on the prepared solid phase was eluted with 10ml of 0.01M nitric acid and measured by inductively coupled plasma-optical emission spectrometry (ICP-OES). The influence of the type and amount of eluent used, pH, sample and eluent flowrates, amount of Cyanex272and the effect of other ions on extraction efficiency were investigated. The limits of detection of the method were 0.3 and 0.1g l-1 for cobalt and nickel, respectively, and provide an enrichment factor of 100. The results obtained on 10 successive extractions and elution cycles revealed relative standard deviations of 1.8 and 1.7% for cobalt and nickel, respectively. The proposed method has been applied to the determination of ultra trace amounts of cobalt and nickel ions in natural and synthetic water samples with satisfactory results.


索引于

  • 中国社会科学院
  • 谷歌学术
  • 打开 J 门
  • 中国知网(CNKI)
  • 引用因子
  • 宇宙IF
  • 电子期刊图书馆
  • 研究期刊索引目录 (DRJI)
  • 秘密搜索引擎实验室
  • ICMJE

查看更多

期刊国际标准号

期刊 h 指数

Flyer