抽象的

Validated HPLC Method for the Determination of Nisoldipine

Shaban A. Abdulla, Manal A. El-Shal and Ali K. Attia


In this study, novel high performance liquid chromatography method which are described as a simple, selective, sensitive, precise and simultaneous analysis of Nisoldipine in bulk and commercial tablet formulation containing Nisoldipine. Good chromatographic separation was achieved using a specific column Agilent ZORBAX Eclipse Plus C18, 4.6 x 250 mm and a mobile phase consisting of methanol, 0.01 M potassium dihydrogen phosphate aqueous solution and 0.1 M Hexane sulphonic acid sodium salt (25 : 65 : 10, v/v) at pH 4.0 using orthophosphoric acid with a flow rate of 1.0 mL/min. The ultraviolet detector was set at a wavelength of 275 nm. Nisoldipine was eluted at 7.43 min. Due to the high precautions taken during the analysis, no extraneous materials were found to interfere. The linear range for Nisoldipine was 5-30 μg/mL. The linearity, precision, accuracy, robustness, limit of detection and limit of quantification of the proposed method were determined. Regression coefficients (r2 ≥ 0.999), recovery (97.2-103.1%), the limit of detection (0.4 μg/mL) and the limit of quantification (1.0 μg/mL) were validated and found to be satisfactory. The proposed method is convenient for quantitative routine analysis and purity control of Nisoldipine in its bulk powder and dosage forms


索引于

  • 中国社会科学院
  • 谷歌学术
  • 打开 J 门
  • 中国知网(CNKI)
  • 引用因子
  • 宇宙IF
  • 米亚尔
  • 秘密搜索引擎实验室
  • 欧洲酒吧
  • 巴塞罗那大学
  • ICMJE

查看更多

期刊国际标准号

期刊 h 指数

Flyer